Parameter | |
Syringe Size | 5ml, 10ml, 20ml, 30ml, 50ml ( optional: 1/2/3mL ) |
Drip Mode |
15 d/ml infusion set 0 d/min~300 d/min
20 d/ml infusion set 0 d/min~400 d/min
60 d/ml infusion set 0 d/min~300 d/min
Note?The minimum resolution dripping speed Settings 1d/min?
|
Accuracy | ?±2%(Note?Under the condition of the benchmark test, using high elastic silicone iv path accuracy after calibration and commissioning?) |
Bubble Detector
|
Ultrasonic detection method?Bubble above 25ul is detected |
Total Infusion Volume
|
0.0ml-9999ml(0.1ml/h increment,1ml/h increment if over 1000ml/h) |
Volume Limit
|
0.0ml-9999ml(0.1ml/h increment,1ml/h increment if over 1000ml/h) |
Occlusion
|
?H??800±200mmHg ?106.7±26.7kPa)
?C??500±100mmHg ?66.7±13.3kPa)
?L??300±100mmHg ?40.7±13.3kPa)
|
History Record | These serial products can maximally store 4500 pieces of history records. The record includes the following information: Date, flow rate, alarm data, total volume delivered, occlusion pressure and syringe brand and serial number? |
Alarm |
Near empty alarm, Empty alarm, Door open alarm, Bubble alarm, limited value reached alarm, AC power failure, Low battery alarm, Battery empty alarm, Remind alarm, Flow regulator or inlet valve closed alarm (optional)
Power Supply Voltage?AC100V?240V?±10%??Power Frequency? 50Hz-60Hz
Built-in Battery?DC15V Ni-MH battery?It can be replaced by lithium ion battery pack
necessary.?
Working time of battery?In case the battery is fully charged, the continuous working time can
be up to 6 hours at the speed of 25ml/h.
Maximum power?30VA?Note?Before the first use of this product, the battery must be charged
while being shut down for more than 12 hours.
|
Working Environment |
Temperature : ?+5?-+40?
Relative humidity : ?20%-90%
Environmental pressure : ?860hPa-1060hPa
|
Transport And Storage Environment |
Temperature : ?-20??+55?
Atmosphere Pressure : ?500hPa?1060hPa.
products should be stored in the relative humidity in the range of 10% ~ 93%?no corrosion gas indoor and ventilated.
Electric Safety Classification Class I, CF type, Built-in battery, Continuous operation equipment
|
Electric Safety Classification | Class I, CF type, Built-in battery, Continuous operation equipment |
Water Proof Level | IPX4 |
Net Weight | 2.2kg(include fastener) |
Dimension | 320mm(W)×132mm?H?×138mm?D? |