Defibrillator:
Defibrillator type Manual Synchronized A synchronized
Output waveform Biphasic technology
Energy accuracy ∠+1%
Energy selection 0.3.5.7.10.20.30.50.100.200.300.360 joule(nominal at a resistance of 50)
Charging time Maximum 8 seconds to 360 Joules
Standard Adult/Pediatric Paddles: Reusable external adult paddles(pediatric paddles integrated)
Patient Monitor:
7”TFT display
ECG:
ECG lead: ⅠⅡⅢ a VR a VL a VF V1-V6
Pulse Rate:
Pulse range: 0-300bpm Pulse accuracy: ±2bpm
NIBP:
Method: automatic oscillation
Working mode: manual/automatic
Measure: adult: 10-250mmHg
Ement: pediatric: 10-200mmHg
Rage: neonatal: 10-135mmHg
TEMP:
Measurement range: 20-45。C
SpO2:
Pulse Oximetry (spO2)
Display: SpO2 value pulse histongram waveform Pulse
Range: 0~99%for adult pediatric and neonatal patients
Probe: standard adult finger clip
AED Mode (Option)
AED Function: Auto analyze and charge x3 with programmable auto energy level selection screen prompts and voice prompts.
Shockable Rhythms: Ventricular fibrillation with amplitude >=200uV ventricular tachycardia with rates >=140bpm and QRS complex wave duration >=140ms.
Charge Control: Control on device front panel press key on paddle.
Prompts: Voice and visual prompts.
Recorder (Printer)(Option)
Paper: 50mm thermal.
Speed: 12.5mm/sec 25mm/sec 50mm/sec. User-selectable 6-second delay.
Printing Method: High-resolution thermal print head.
Print-out Modes: Manual or automatic user-configurable.
On/Off Control: Front panel and paddle.
Automatic Function: 9-seconds recording initiated by alarm activation or defibrillator charge or defibrillator discharge.