Standard parameter & configuration
Defi8: ECG .PR .DEFIBRILLATOR
Options:
Recorder
Dimensions: L320 x W205 x H410
Weight: 9.8kg
Technical specifications
Defibrillator
Defibrillator type Manual Synchronized Asynchronized
Out put waveform Biphasic technology
Energy accuracy <+1%
Energy selection 0.3.5.7.10.20.30.50.100.200.300.360 joule (nominal at a resistance of 50)
Charging time Maximum 8 seconds to 360 Joules
Standard Adult/Pediatric Paddles
Reusable external adult paddles (pediatric paddles integrated)
Power requirements
AC Power supply 110V 60 Hz /220V50 Hz
Vehicle Voltage DC 12V
Internal Battery Backup
Battery type Rechargeable lead-acid battery/ lithium battery(optional)
Charging time Time-Minimum of 4 hours for full charging
Battery charging During AC Power Supply operation of the device.
Charging /Low Battery Indication Red LED
Battery Capacity 100% of the battery will provide 120 mins of monitoring and 30 discharges at 360 joules.
Display
LED Dimensions 7’diagonal {155mm*88mm}
Type High resolution true-color graphic LED
Resolution 480X234
ECG
ECG lead: I II III AVR AVL AVF V1 V6 GND
S-T segment analysis: -2.0/2.0mv
Arrhythmia analysis: yes
Protection: With stand 4000vAc/50Hz voltage in isolation and work against eletrosurgical interfere and defibrillation
Pulse range
Pulse range: 0-255bpm
Pulse Oximetry (spO2) (Option)
Display: SpO2 value pulse histongram waveform Pulse
Range: 0~99%for adult pediatric and neonatal patients
Probe: standard adult finger clip
AED Mode (Option)
AED Function: Auto analyze and charge x3 with programmable auto energy level selection screen prompts and voice prompts.
Shockable Rhythms: Ventricular fibrillation with amplitude >=200uV ventricular tachycardia with rates >=140bpm and QRS complex wave duration >=140ms.
Charge Control: Control on device front panel press key on paddle.
Prompts: Voice and visual prompts.
Recorder (Printer) (Option)
Paper: 50mm thermal.
Speed: 12.5mm/sec 25mm/sec 50mm/sec. User-selectable 6-second delay.
Printing Method: High-resolution thermal print head.
Print-out Modes: Manual or automatic user-configurable.
On/Off Control: Front panel and paddle.
Automatic Function: 9-seconds recording initiated by alarm activation or defibrillator charge or defibrillator discharge.
AED Mode (Option)
AED Function: Auto analyze and charge x3 with programmable auto energy level selection screen prompts and voice prompts.
Shockable Rhythms: Ventricular fibrillation with amplitude >=200uV ventricular tachycardia with rates >=140bpm and QRS complex wave duration >=140ms.
Charge Control: Control on device front panel press key on paddle.
Prompts: Voice and visual prompts.
Recorder (Printer)(Option)
Paper: 50mm thermal.
Speed: 12.5mm/sec 25mm/sec 50mm/sec. User-selectable 6-second delay.
Printing Method: High-resolution thermal print head.
Print-out Modes: Manual or automatic user-configurable.
On/Off Control: Front panel and paddle.
Automatic Function: 9-seconds recording initiated by alarm activation or defibrillator charge or defibrillator discharge.